| Standard | Description | Application |
| UL 60601-1 | Electrical safety measure required for all devices that can potentially come in contact with patients and staff. | Medical equipment-patient care vicinity |
| UL 1363A | Basic safety standard for power strips integrated as components of a larger system where they supply power to equipment in a movable assembly that is rack, table or pedestal-mounted. | Mobile applications inside and outside the patient care vicinity |
| UL 1363 | Basic safety standard for power strips that stand alone and are NOT designed as components of a larger system. | Lab applications and administrative areas |
| UL 1449 | Surge protector safety standard for units used outside of patient care vicinity. | Administrative areas and business offices |
If so, your power strip must meet UL 60601-1, the standard for electrical safety required for all devices that potentially come in contact with patients and staff.

If so, your power strip must meet UL 1363A, the standard required for power strips mounted as components of movable equipment assemblies.

Are they ever moved from one room to another? If so, your surge protector must meet UL 1449 and have hospital-grade outlets. This ensures safety when strips are transferred between locations.


UL60601-1 compliance designed for use anywhere within patient-care and treatment rooms.
Ideal for powering medical and computer equipment.

Perfect for mobile applications inside and outside patient-care and treatment rooms. Designed for powering movable plug-connected components mounted to IV poles, mobile carts, crash carts and mobile medical workstations.

Extra-long 25 foot power cord designed for mobile applications inside and outside operating, patient-care and treatment rooms. Ideal for powering movable plug-connected components mounted to IV poles, mobile carts, crash carts and mobile medical workstations.